Seedlabs
Source research

Assessment of the reproducibility and precision of milling and 3D printing surgical guides

Sueli Mukai, Eduardo Mukai, José Arnaldo Santos-Junior, Jamil Awad Shibli, Marcelo Faveri, Gabriela Giro · 2021 · 95 citationsRead the paper

BACKGROUND: Technology advancement has rising in the past decade and brought several innovations and improvements. In dentistry, this advances provided more comfortable and quick procedures to both the patient and the dental surgeon, generating less predictability in the final result. Several techniques has been developed for the preparation of surgical guides aiming at the optimization of surgical procedures. The present study aimed to evaluate the reproducibility and precision of two types of surgical guides obtained using 3D printing and milling methods. METHODS: A virtual model was developed that allowed the virtual design of milled (n = 10) or 3D printed (n = 10) surgical guides. The surgical guides were digitally oriented and overlapped on the virtual model. For the milling guides, the Sirona Dentsply system was used, while the 3D printing guides were produced using EnvisionTEC's Perfactory P4K Life Series 3D printer and E-Guide Tint, a biocompatible Class I certified material. The precision and trueness of each group during overlap were assessed. The data were analyzed with GraphPad software using the Kolmogorov-Smirnov test for normality and Student's t test for the variables. RESULTS: The Kolmogorov-Smirnov test showed a normal distribution of the data. Comparisons between groups showed no statistically significant differences for trueness (p = 0.529) or precision (p = 0.3021). However, a significant difference was observed in the standard deviation of mismatches regarding accuracy from the master model (p < 0.0001). CONCLUSIONS: Within the limits of this study, surgical guides fabricated by milling or prototyped processes achieved similar results.

What the full text shows

Shown in the lab
  • No statistically significant difference in trueness between milled and 3D printed guides. — p = 0.529 (Comparison of 10 milled vs 10 3D printed guides)
  • No statistically significant difference in precision between milled and 3D printed guides. — p = 0.3021 (Comparison of 10 milled vs 10 3D printed guides)
  • 3D printing guides showed significantly higher variation in accuracy from the master model. — p < 0.0001 (standard deviation of mismatches) (Comparison of 10 milled vs 10 3D printed guides)
Setup
20 surgical guides total (10 milled PMMA, 10 3D printed E-Guide Tint) based on a single partially edentulous Kennedy class IV model.
How
Virtual surgical guides were designed and fabricated using either a milling machine (Sirona Dentsply) or a DLP 3D printer (EnvisionTEC). The resulting guides were digitized and superimposed on the virtual master model using Optocat software to measure mismatch, precision, and trueness.
Limitations
In vitro study using a virtual model rather than clinical patients. · Small sample size (n=10 per group). · Does not measure actual implant placement deviation in bone.

Summarized by Gemma 4 from the open-access full text · 3 Oct 2026. Read it