Seedlabs
Source research

Clinical validation of an automatic atlas‐based segmentation tool for male pelvis CT images

Marta Casati, Stefano Piffer, Silvia Calusi, L. Marrazzo, Gabriele Simontacchi, Vanessa Di Cataldo et al. · 2022 · 17 citationsRead the paper

All 12 authors

Daniela Greto, Isacco Desideri, Marco Vernaleone, Giulio Francolini, Lorenzo Livi, S. Pallotta

PURPOSE: This retrospective work aims to evaluate the possible impact on intra- and inter-observer variability, contouring time, and contour accuracy of introducing a pelvis computed tomography (CT) auto-segmentation tool in radiotherapy planning workflow. METHODS: Tests were carried out on five structures (bladder, rectum, pelvic lymph-nodes, and femoral heads) of six previously treated subjects, enrolling five radiation oncologists (ROs) to manually re-contour and edit auto-contours generated with a male pelvis CT atlas created with the commercial software MIM MAESTRO. The ROs first delineated manual contours (M). Then they modified the auto-contours, producing automatic-modified (AM) contours. The procedure was repeated to evaluate intra-observer variability, producing M1, M2, AM1, and AM2 contour sets (each comprising 5 structures × 6 test patients × 5 ROs = 150 contours), for a total of 600 contours. Potential time savings was evaluated by comparing contouring and editing times. Structure contours were compared to a reference standard by means of Dice similarity coefficient (DSC) and mean distance to agreement (MDA), to assess intra- and inter-observer variability. To exclude any automation bias, ROs evaluated both M and AM sets as "clinically acceptable" or "to be corrected" in a blind test. RESULTS: Comparing AM to M sets, a significant reduction of both inter-observer variability (p < 0.001) and contouring time (-45% whole pelvis, p < 0.001) was obtained. Intra-observer variability reduction was significant only for bladder and femoral heads (p < 0.001). The statistical test showed no significant bias. CONCLUSION: Our atlas-based workflow proved to be effective for clinical practice as it can improve contour reproducibility and generate time savings. Based on these findings, institutions are encouraged to implement their auto-segmentation method.

What the full text shows

Working prototype
  • Reduction in whole pelvis contouring time when using auto-segmentation followed by manual editing. — -45% vs. manual contouring (Average across 5 ROs and 6 test subjects)
  • Reduction in inter-observer variability for all regions of interest (bladder, rectum, pelvic lymph-nodes, femoral heads). — p < 0.001 for DSC and MDA (Compared to manual contouring)
  • Reduction in intra-observer variability for specific structures. — p < 0.001 for bladder and femoral heads (Compared to manual contouring)
  • No statistically significant reduction in intra-observer variability for specific structures. — No significant difference (Rectum and pelvic lymph-nodes)
  • Reduction in the 95th percentile of Mean Distance to Agreement (MDA) for specific structures. — Bladder: 1.0 to 0.7 mm; PLN: 3.2 to 1.5 mm; Femoral heads: 1.1 to 0.4 mm and 1.0 to 0.6 mm (M to AM contours)
Setup
6 test subjects, 5 radiation oncologists, 5 structures (bladder, rectum, pelvic lymph-nodes, femoral heads), 600 total contours.
How
A multi-subject atlas-based segmentation tool (MIM MAESTRO) was used to generate contours for five structures in male pelvis CTs. Five radiation oncologists performed manual contouring (M) and editing of auto-contours (AM) twice each to evaluate intra- and inter-observer variability and time savings.
Limitations
Small sample size (6 patients, 5 ROs). · Retrospective study design. · Lack of intra-observer reduction for rectum and PLN attributed to intrinsic delineation uncertainty and poor contrast.

Summarized by Gemma 4 from the open-access full text · 13 Nov 2026. Read it