Ultra-highly sensitive HCl-LITES sensor based on a low-frequency quartz tuning fork and a fiber-coupled multi-pass cell
Shunda Qiao, Angelo Sampaolo, Pietro Patimisco, Vincenzo Spagnolo, Yufei Ma · 2022 · 108 citationsRead the paper
In this paper, an ultra-highly sensitive light-induced thermoelastic spectroscopy (LITES) based hydrogen chloride (HCl) sensor, exploiting a custom low-frequency quartz tuning fork (QTF) and a fiber-coupled multi-pass cell (MPC) with optical length of 40 m, was demonstrated. A low resonant frequency of 2.89 kHz of QTF is advantageous to produce a long energy accumulation time in LITES. Furthermore, the use of an MPC with the fiber-coupled structure not only avoids the difficulty in optical alignment but also enhances the system robustness. A distributed feedback (DFB) diode laser emitting at 1.74 µm was used as the excitation source. Under the same operating conditions, the using of low-frequency QTF provided a ~2 times signal improvement compared to that achieved using a standard 32 kHz QTF. At an integration time of 200 ms, a minimum detection limit (MDL) of 148 ppb was achieved. The reported sensor also shows an excellent linear response to HCl gas concentration in the investigated range.
What the full text shows
Working prototype- Low-frequency QTF provided a signal improvement over standard 32 kHz QTF. — ~2 times higher signal (500 ppm HCl:N2 gas mixture)
- Minimum detection limit (MDL) achieved with low-frequency QTF. — 148 ppb (1σ) and 444 ppb (3σ) (Integration time of 200 ms)
- Linear response of the sensor to HCl gas concentration. — R-squared of 0.99 (Investigated concentration range)
- Reproducible quantification concentration fluctuation. — 0.45 ppm (Based on 3σ noise and linear response)
- Setup
- Custom QTF (2.89 kHz) vs commercial QTF (32 kHz), fiber-coupled MPC with 40 m optical length, DFB diode laser (1.74 µm), HCl:N2 gas mixtures.
- How
- The researchers developed a Light-Induced Thermoelastic Spectroscopy (LITES) sensor using a custom low-frequency quartz tuning fork (QTF) and a fiber-coupled multi-pass cell (MPC). They employed a DFB diode laser at 1.74 µm and used wavelength modulation spectroscopy (WMS) with 2nd harmonic detection to enhance sensitivity.
Summarized by Gemma 4 from the open-access full text · 24 Oct 2026. Read it