Highly sensitive HF detection based on absorption enhanced light-induced thermoelastic spectroscopy with a quartz tuning fork of receive and shallow neural network fitting
Xiaonan Liu, Shunda Qiao, Guowei Han, Jinxing Liang, Yufei Ma · 2022 · 98 citationsRead the paper
Due to its advantages of non-contact measurement and high sensitivity, light-induced thermoelastic spectroscopy (LITES) is one of the most promising methods for corrosive gas detection. In this manuscript, a highly sensitive hydrogen fluoride (HF) sensor based on LITES technique is reported for the first time. With simple structure and strong robustness, a shallow neural network (SNN) fitting algorithm is introduced into the field of spectroscopy data processing to achieve denoising. This algorithm provides an end-to-end approach that takes in the raw input data without any pre-processing and extracts features automatically. A continuous wave (CW) distributed feedback diode (DFB) laser with an emission wavelength of 1.27 µm was used as the excitation source. A Herriott multi-pass cell (MPC) with an optical length of 10.1 m was selected to enhance the laser absorption. A quartz tuning fork (QTF) with resonance frequency of 32,767.52 Hz was adopted as the thermoelastic detector. An Allan variance analysis was performed to demonstrate the system stability. When the integration time was 110 s, the minimum detection limit (MDL) was found to be 71 ppb. After the SNN fitting algorithm was used, the signal-to-noise ratio (SNR) of the HF-LITES sensor was improved by a factor of 2.0, which verified the effectiveness of this fitting algorithm for spectroscopy data processing.
What the full text shows
Working prototype- Minimum detection limit (MDL) for HF gas achieved. — 71 ppb at 110 s integration time (Using LITES with a Herriott multi-pass cell and QTF detector)
- Improvement in signal-to-noise ratio (SNR) via SNN fitting algorithm. — 2.0x increase (Applied to raw spectroscopy data for denoising)
- Minimum detection limit (MDL) at short integration time. — 370 ppb at 100 ms integration time (Based on Allan variance analysis)
- Normalized noise equivalent absorption (NNEA) of the sensor. — 2.28 × 10^-6 cm^-1 W/Hz^1/2 (HF-LITES sensor setup)
- Linearity of HF-LITES signal versus gas concentration. — R-square ~ 0.98 (HF gas mixtures)
- Setup
- Herriott multi-pass cell (10.1 m optical length), QTF (resonance frequency 32,767.52 Hz), 1.27 µm CW-DFB laser, and a lock-in amplifier.
- How
- The researchers used a 1.27 µm CW-DFB laser and a Herriott multi-pass cell to excite HF gas, detecting the resulting thermoelastic effect using a quartz tuning fork (QTF) via wavelength modulation spectroscopy (WMS). A shallow neural network (SNN) with one hidden layer and a sigmoid activation function was implemented to denoise the raw spectroscopy data.
- Limitations
- Strong adsorption of HF gas on surfaces leads to slower filling times (220 s) and slightly reduced linearity (R^2 ~ 0.98). · System stability is dominated by white noise only up to an integration time of 135 s.
Summarized by Gemma 4 from the open-access full text · 24 Oct 2026. Read it