Comparison of the cytotoxicity of 3D-printed aligners using different post-curing procedures: an in vitro study

Alessandra Campobasso, Anastasia Ariano, Giovanni Battista, Posa Francesca, Marco Migliorati, Sara Drago et al. · 2023 · 47 citationsRead the paper

All 8 authors

Eleonora Lo Muzio, Mori Giorgio

Objective Three-dimensional (3D) printing technology represents a novel method for manufacturing aligners. The aim of the present study was to assess the in-vitro cytotoxicity of 3D-printed aligners using different post-polymerisation conditions. Materials Aligners were printed using the same 3D-print resin (TC-85DAC, Graphy, Seoul, Korea) and printer (AccuFab-L4D, Shining 3D Tech. Co., Hangzhou, China), followed by different post-curing procedures. Six aligners were post-polymerised for 14 min using the Tera Harz Cure and a nitrogen generator curing machine (THC2, Graphy, Seoul, Korea) (P1). A further six aligners were post-cured for 30 min on each side using the Form Cure machine (FormLabs Inc, Somerville, USA) (P2). The aligners were cut into smaller specimens (2 mm×2 mm) and sterilised at 121°C. The specimens were placed in 96-well plates containing Dulbecco’s Modified Eagle’s Medium (DMEM) at 37° for 7 or 14 days. The viability of MC3T3E-1 pre-osteoblasts cultured with DMEM was evaluated using the 3-(4,5-dimethylthiazol-2-yl)-2,5-diphenyltetrazolium bromide (MTT) assay. The optical density of each cell culture was measured to assess cell viability, following which the data were statistically analysed using two-way and one-way ANOVA (α = 0.05). Results The comparison of cytotoxicity revealed statistically significant differences between post-curing procedures and MTT timings ( P < 0.001). After 7 and 14 days, the cell viability of P2 was significantly reduced compared to P1 and the control groups ( P < 0.001), while P1 showed no significant differences compared to the controls. Overall, P2 post-curing exhibited moderate cytotoxicity, while P1 post-polymerisation was highly biocompatible. Conclusions Different post-curing procedures may affect the in-vitro cytotoxicity of 3D-printed aligners. Clinicians should adhere to the manufacturer’s recommendations when using 3D-print resin.

What the full text shows

Shown in the lab
  • Post-curing with a nitrogen generator (P1) showed high biocompatibility with no significant difference from control groups. — 107.12% ± 17.47% viability (7 days) and 106.74% ± 18.41% viability (14 days) (TC-85DAC resin, THC2 curing machine with nitrogen generator, 14 min curing)
  • Post-curing without a nitrogen generator (P2) exhibited moderate cytotoxicity, significantly reducing cell viability. — 59.79% ± 10.06% viability (7 days) and 47.09% ± 20.62% viability (14 days) (TC-85DAC resin, Form Cure machine, 60 min total curing (30 min per side))
Setup
12 aligners printed using AccuFab-L4D printer and TC-85DAC resin; MC3T3-E1 pre-osteoblast mouse calvaria cells in 96-well plates.
How
3D-printed aligners were subjected to two different post-curing protocols and then cut into 2mm x 2mm specimens. Cell viability was measured using an MTT assay on MC3T3-E1 pre-osteoblasts over 7 and 14 days.
Limitations
In vitro study using pre-osteoblasts rather than in vivo oral tissue · Small sample size (N=12 aligners)