Accuracy and safety of real-time navigation during insertion of temporary anchorage devices - A randomized clinical trial
Yi-Cheng Chen, Hung‐Ying Lin, Sang‐Heng Kok, Hao-Hueng Chang, Chun-Pin Lin · 2025 · 2 citationsRead the paper
Background/purpose: Temporary anchorage device (TAD) is one of the applications of dental implants that facilitate orthodontic treatment. However, there are possible complications in TAD surgery with the freehand technique. Navigation has been used in dental implantation with acceptable results. There has been no published study on the use of real-time navigation for TAD insertion. Materials and methods: This was a split-mouth design randomized clinical trial in which the control group used a surgical guide (N = 16) for TAD insertion surgery and the experimental group used navigation for surgery (N = 16). We also performed a negative control by retrospective database survey to evaluate the freehand technique (N = 29). By comparing the platform center deviation and angular deviation of the control group, experimental group, and negative control, we can draw conclusions about the accuracy and safety of TAD insertion surgery using each method. Results: < 0.05), indicating that the surgical guide provided better position control. There was no statistical difference in terms of angular deviation. Conclusion: The clinical use of real-time navigation in the insertion of TAD offers no more significant advantages than the freehand technique in terms of position and angular control. No immediate complications requiring implant removal occurred in this study. The combined technique has shown a trend in accuracy and safety, but a larger number of samples are still needed for statistical analysis to draw a definitive conclusion.
What the full text shows
Tested in real conditions- Surgical guides provided significantly better position control than real-time navigation. — Mean platform center deviation: 1.17 ± 0.77 mm (guide) vs. 2.99 ± 1.75 mm (navigation); P = 0.010 (Comparison between control and experimental groups)
- Real-time navigation showed no statistically significant difference in position control compared to the freehand technique. — Mean platform center deviation: 2.99 ± 1.75 mm (navigation) vs. 3.10 ± 1.91 mm (freehand); P = 1.000 (Comparison between experimental and negative control groups)
- No statistically significant difference in angular deviation was found between surgical guides and real-time navigation. — Mean angular deviation: 12.43 ± 8.32° (guide) vs. 12.49 ± 4.48° (navigation); P = 0.291 (Comparison between control and experimental groups)
- No statistically significant difference in angular deviation was found between real-time navigation and the freehand technique. — Mean angular deviation: 12.49 ± 4.48° (navigation) vs. 15.64 ± 5.94° (freehand); P = 0.170 (Comparison between experimental and negative control groups)
- Real-time navigation was safe with no immediate complications requiring implant removal. — 0 complications (Experimental group)
- Setup
- 61 TADs total; 16 samples in control group, 16 in experimental group, and 29 in negative control group.
- How
- A split-mouth randomized clinical trial compared TAD insertion using a 3D-printed surgical guide (control) and the IRIS® real-time navigation system (experimental), with a retrospective freehand group (negative control). Accuracy was measured by superimposing postoperative CBCT scans with preoperative plans to calculate platform center and angular deviations.
- Limitations
- Small sample size · Signal noise in CBCT caused by orthodontic brackets affecting 3D model accuracy · Potential instability of the navigation tracker due to the use of utility wax · Operator familiarity with the navigation system (human error)
Summarized by Gemma 4 from the open-access full text · 1 Nov 2026. Read it