Accuracy of guided insertion of orthodontic temporary anchorage devices comparing two different 3D printed surgical guide designs: a randomized controlled trial
Lea Hoffmann, Tamara Katharina Kakoschke, Alexander Keller, Linus Hötzel, Uwe Baumert, Sven Otto et al. · 2025 · 4 citationsRead the paper
All 8 authors
Andrea Wichelhaus, Hisham Sabbagh
Temporary anchorage devices (TADs) provide maximum anchorage in orthodontic therapy. Surgical guides can be used to accurately achieve pre-planned TAD positions. The objective of this study was to assess the accuracy of guided TAD insertion in the anterior palate, comparing two 3D printed surgical guide designs. Patients were randomized in two groups. TADs were inserted using either a surgical guide with a skeletonized design or full arch design. The primary outcome was translational and rotational deviations from the virtually planned TAD positions. Secondary outcomes included the dependence of deviations on age and gender and TAD success rate after a follow-up period of three months. In total 40 patients were included in this RCT and 80 TADs were inserted. During the follow-up period, no TAD loss was observed. The skeletonized design showed significantly higher deviations between the planned TAD position and the actual TAD position in all, but two spatial dimensions (translational and rotational) compared to the full arch design. The right TAD showed higher deviations compared to the left TAD for the skeletonized design. The full arch design showed significantly higher accuracy compared to the skeletonized design. During virtual planning of TAD positions, a safety distance should be considered to account for possible deviations using 3D printed surgical guides for TAD insertion in the anterior palate.
What the full text shows
Tested in real conditions- Full arch design demonstrated significantly higher accuracy than skeletonized design across nearly all spatial dimensions. (TAD insertion in the anterior palate)
- Mean angular deviation was lower for the full arch design compared to the skeletonized design. — 5.55° vs 6.44° (Comparison of planned vs actual TAD position)
- Mean distance deviation at the tip was lower for the full arch design. — 1.47 mm vs 2.00 mm (Comparison of planned vs actual TAD position)
- Mean distance deviation at the top was lower for the full arch design. — 0.73 mm vs 1.01 mm (Comparison of planned vs actual TAD position)
- TAD success rate was absolute for both guide designs. — 100% success (3-month follow-up period after orthodontic loading)
- Skeletonized design showed positive trends for distance tip, distance top, and angle deviations relative to age. — R² = 0.230 (tip), 0.093 (top), 0.093 (angle) (Skeletonized design group)
- Setup
- 40 patients (22 female, 18 male; mean age 19.09 ± 9.74 years), 80 TADs (2.3 x 9.0 mm) inserted in the anterior palate.
- How
- A randomized controlled trial compared two 3D-printed SLA surgical guides (full arch vs. skeletonized) for TAD insertion. Accuracy was measured by matching post-insertion intraoral scans of scan abutments to virtual planning models using an ICP algorithm.
- Limitations
- Deviations were noted as comparatively high compared to some other clinical studies. · The clinical protocol of always inserting the left TAD first may have influenced the results for the skeletonized design. · Small sample size (40 patients).
Summarized by Gemma 4 from the open-access full text · 8 Oct 2026. Read it